Development and investigation scintillating optical fiber-based radiometer for measuring the activities of 63Ni isotope radiation sources
Автор: Novikov Sergey, Berintsev Alexey, Svetukhin Vyacheslav, Alexeyev Alexander, Chertoriysky Alexey, Kuznetsov Rostislav, Andreychuk Nikolay
Журнал: Известия Самарского научного центра Российской академии наук @izvestiya-ssc
Рубрика: Физика и электроника
Статья в выпуске: 6-1 т.16, 2014 года.
Бесплатный доступ
An experimental sample of radiometer based on the «Saint-Gobain Crystals» BSF-60 scintillating optical fiber has been developed aimed at measuring the activities of flat ionizing radiation sources based on 63Ni isotope. An experimental study of the dosimeter sensitivity has been implemented within the activities range of the radiation source up to 98 mCi demonstrating that the count rate at the radiation meter output is proportional to the relative radiation activity of the ionizing radiation source based on 63Ni isotope.
Scintillation fiber, source of ionizing radiation, dosimetry system
Короткий адрес: https://sciup.org/148203539
IDR: 148203539