Improving X-ray microtomography resolution

Автор: Shtyrlyaeva A.A., Zhuravlev A.V.

Журнал: Вестник геонаук @vestnik-geo

Рубрика: Научные статьи

Статья в выпуске: 6 (258), 2016 года.

Бесплатный доступ

We suggest the method to increase the radiodensity resolution of the industrial microtomographes (SkyScan). The method is based on the filter with a narrow transmission spectrum (60-90 keV with 100-130 kV source voltage). The filter is composed of lead-containing glass 2.14 mm thick.

X-ray microtomography, source voltage, filter, spatial resolution, radiodensity, core sample

Короткий адрес: https://sciup.org/149128677

IDR: 149128677   |   DOI: 10.19110/2221-1381-2016-6-24-27