Diagnostic value of the portable urine analyzer "ETTA AMP-01" as a tool for self-monitoring in mHealth and screening in primary care
Автор: Shaderkin I.A., Vladzymyrskyy A.V., Tsoy A.A., Voytko D.A., Prosyannikov M.Y., Zelenskii M.M.
Журнал: Экспериментальная и клиническая урология @ecuro
Рубрика: Экспериментальная урология
Статья в выпуске: 4, 2015 года.
Бесплатный доступ
Urinalysis is the most commonly used method of assessing and monitoring of the patient in urological practice. Objective: to estimate the diagnostic value of the portable urine analyzer "ETTA AMP-01" for subsequent use as a tool for patient-centered urology (mHealth) with proven efficacy. Materials and Methods. Results of 109 patients urine analysis performed with a portable urine analyzer ETTA AMP-01 and standard tools of laboratory examination were compared. Protein, urobilinogen, erythrocytes, leukocytes, ketone bodies, glucose and nitrite were identified. We assessed diagnostic consideration, sensitivity and specificity; all results are staistically reliable. Corresponding rates in comparison to standard laboratory methods are as follows: protein - 0.69, 46.77%. 100.00%; urobilinogen - 0.98, 40.00% 100.00%; erythrocytes - 0.91, 74% 93.55%; leukocytes - 0.85, 67.35%, 100.00%; ketone bodies - 0.98, 71.43%, 100.00%; glucose - 0.99, 66.67%, 100.00%; nitrites - 0.96 100.00%, 95.79%. Diagnostic accuracy: 0,69-0,99 (> 0.9 for 5 out of 7 indicators). The specificity of the analyzer: 93,6100,0% (average - 0.99 ± 0.03). What characterizes the analyzer to be more specific than sensitive diagnostic device that is common to many screening tools. Conclusion. Studied method is a specific diagnostic test with high predictive value and diagnostic accuracy.
Urine analysis, urine analyzer, test strips, screening, mhealth
Короткий адрес: https://sciup.org/142188063
IDR: 142188063