Investigation of characteristics of a gas-dynamic interface bandwidth time-of-flight mass-spectrometer MX 5311

Автор: Gavrik M.A., Kornev D.I., Muradymov M.Z.

Журнал: Научное приборостроение @nauchnoe-priborostroenie

Рубрика: Масс-спектрометрия для биотехнологии

Статья в выпуске: 4 т.20, 2010 года.

Бесплатный доступ

In the work the gas-dynamic interface time-of-flight mass spectrometer MX 5311 with orthogonal input of ions and the ESI ion source is described. The results of the research of gas-dynamic interface bandwidth TOFMS MX 5311 are presented. It is shown that the complete loss of ion current during the passage of gas-dynamic interface from the field ionization chamber to the mass analyzer is less than 6 orders of magnitude.

Mass-spectrometry, tof ms, gas-dynamic interface, radiofrequency quadrupole, electrospray, parameters optimization, bandwidth interface