Clinical performance of system for ultrasonic cortex aspiration during phacoemulsification

Автор: Aznabaev В.M., Dibaev T.I., Mukhamadeev T.R., Idrisova G.M.

Журнал: Саратовский научно-медицинский журнал @ssmj

Рубрика: Дерматовенерология

Статья в выпуске: 4 т.14, 2018 года.

Бесплатный доступ

Purpose: to study the performance and safety of the use of system for cortex aspiration with the possibility of additional dosed low power ultrasound during phacoemulsification. Material and Methods. The comparative evaluation of developed system for cortex aspiration with the possibility of additional dosed low power ultrasound (main group, n=36, ophthalmological surgical platform Optimed Profi) and the traditional system for cortex aspiration (control group, n=40, ophthalmological surgical platform Centurion Vision System) were performed. Results. Using the system for cortex aspiration decreases mean time of cortex aspiration. In the main group the mean time of cortex aspiration was 72.6±24.3 s, in the control group 93.9±48.0 s (p

Еще

Cataract, ultrasonic phacoemulsification, irrigation and aspiration of cortex, ultrasonic aspiration, time of aspiration