Accuracy assessment of body composition measurement with KLU

Автор: Piskaev A.A., Korableva Yu.B., Epishev V.V., Nmeth Z., Balakhonova L.D.

Журнал: Человек. Спорт. Медицина @hsm-susu

Рубрика: Физиология

Статья в выпуске: 1 т.22, 2022 года.

Бесплатный доступ

Aim. The paper aims to assess the accuracy of body composition measurement by KLU. Materials and methods. The study involved 32 subjects (males n = 11; females n = 21) aged 19-35 years (25.51 ± 5.82). The following equipment were used for the purpose of the study: height meter, Tanita BC-418 MA body composition analyser, KLU body composition tracker. Results. Significant differences in body fat (more than 20%) were found in 17 subjects. In 6 subjects, this difference was more than 40%. Therefore, in 53.12% of subjects the results obtained with different equipment did not coincide (63.63% - males; 47.61% - females). Better results were reported when measuring body water using the abovementioned equipment. There were no differences of more than 20%, while a 10%-difference was found in 15.62% of subjects. Conclusion. The difference of more than 20% reported in 53.12 % of subjects does not allow to consider KLU as an accurate device for body composition measurement or an accurate individual device. Therefore, its personal use cannot be a reliable method of body composition measurement.

Еще

Body composition tracker, bioimpedance, measurement accuracy

Короткий адрес: https://sciup.org/147237178

IDR: 147237178

Статья научная