Establishment of the calibration line by DNA fragment analysis

Автор: Belov Yu. V., Leontyev I.A., Panchuk V.V., Petrov A.I., Kurochkin V.E.

Журнал: Научное приборостроение @nauchnoe-priborostroenie

Рубрика: Использование и методики высокотехнологичных измерений

Статья в выпуске: 3 т.23, 2013 года.

Бесплатный доступ

It’s proposed to use an approximation of the tabulated dependence of the peaks position of the internal standard on the time by a polynomial of the fourth degree as an analytic function. This analytic function permits to plot a calibration line to determine the optimal length of the analyzed DNA fragments within the calibration range and beyond its range.

Dna, genetic analyzer, fragment analysis

Короткий адрес: https://sciup.org/14264871

IDR: 14264871

Статья научная