New software to standardize the assessment of embolization of cerebral arteriovenous malformations

Автор: Sergeyev G.S., Orlov K.Yu., Krivoshapkin A.L., Panarin V.A., Kurbatov V.P., Amelin M.E., Kalneus L.E., Chebykin D.V.

Журнал: Патология кровообращения и кардиохирургия @journal-meshalkin

Рубрика: Ангионеврология и нейрохирургия

Статья в выпуске: 1 т.17, 2013 года.

Бесплатный доступ

Staged embolization of high-grade Spetzler-Martin AVMs gains popularity. The assessment of AVM devascularization degree depends on a specialist's personal experience and is very subjective. The goal of our study was to develop software for accurate and objective calculation of AVM devascularization percentage. 34 patients underwent a partial AVM embolization procedure using both adhesive and non-adhesive embolic material. Their pre- and postoperative DICOM angiographic images were assessed by both 4 skilled interventional radiologists and by our new software. We compared the evaluations of AVM devascularization degree made by the software and the specialists. 54 tests were performed. The software was based on a LabVIEW tm platform. The degree of AVM devascularization was estimated in percent values. Mean difference of values was 14,8% ± 1,6 (range 0%-40%) between human estimations and 3,4% ± 0,4 (range 0,7-9,2%) between computer calculations. Reliability of the samples was estimated by Kolmogorov-Smirnov criterion. Critical value of the criterion was λ2 = 4,89, which corresponds to p value

Еще

Arteriovenous malformation, medical software, degree of avm devascularization

Короткий адрес: https://sciup.org/142140519

IDR: 142140519

Статья научная