Design of an elliptical coil for a specialized magnetic resonance scanner with a 0.4 Tesla field
Автор: Bayazitov Alfis Albertovitch, Fattakhov Ya. V., Khundiryakov V.E.
Журнал: Научное приборостроение @nauchnoe-priborostroenie
Рубрика: Работы с конференции
Статья в выпуске: 1 т.29, 2019 года.
Бесплатный доступ
The work is devoted to mathematical calculation, development of designs measurement of radio frequency characteristics of the coil for a specialized magnetic resonance scanner. The results of mathematical modeling of the field homogeneity, for the receiving coil, are presented depending on its design features. A comparative analysis of the results of two types of contours are made: cylindrical and elliptical. Based on the results of the simulation, experimental samples of the receiving coil are manufactured and the results of measuring their radio-frequency characteristics are presented.
Mri, calculation of magnetic field distribution, sensor "brush", the receiving sensor, gauge brush
Короткий адрес: https://sciup.org/142218198
IDR: 142218198 | DOI: 10.18358/np-29-1-i9298