In vitro implantation of a new transcatheter prosthesis in mitral valve position
Автор: Bogachev-prokophiev Alexander V., Zhuravleva Irina Yu., Sharifulin Ravil M., Zheleznev Sergey I., Demidov Denis P., Kliver Evgeny E., Karaskov Alexander M.
Журнал: Патология кровообращения и кардиохирургия @journal-meshalkin
Рубрика: Приобретенные пороки сердца
Статья в выпуске: 1 т.22, 2018 года.
Бесплатный доступ
Aim. The study was designed to focus on the evaluation of in vitro implantation of a new transcatheter mitral bioprosthesis. Methods. A prototype of the first domestic self-expanding transcatheter bioprosthesis “Solertis” was tested by implanting it in the position of a native mitral valve of isolated pig heart by using transatrial access. Results. A correct orientation of the atrial and ventricular elements of the bioprosthesis and reliable circular coverage/adaptation of the annular part of the stent, which provide stable fixation of the prosthesis in a mitral position without left ventricular outflow tract obstruction, were determined. The function of the prosthesis during saline load test was adequate and without paraprosthetic regurgitation. Conclusion. The prototype of “Solertis” self-expanding transcatheter bioprosthesis demonstrated its efficiency for the native mitral valve replacement in vitro. The study results allow for proceeding to the next stage of predinical testing-in vivo investigation.
Mitral valve, bioprosthesis, transcatheter mitral valve implantation
Короткий адрес: https://sciup.org/142230645
IDR: 142230645 | DOI: 10.21688/1681-3472-2018-1-22-28