Effects of biocompatible extracorporeal bypass circuit coating Bioline in cardiac surgery patients

Бесплатный доступ

Biocompatibility of heparin-based coating for extracorporeal circuits, the Bioline, was evaluated in cardiac surgery setting. Thirty six CABG pts, aged from 30 to 69 yrs, were randomly perfused with a circuit containing Bioline-coated Quadrox oxygenator (Bioline, n=18) or with a similar non-coated circuit (control, n=18). Leukocyte and platelet counts, levels of tumor necrosis factor, interleukin-6 and fibrinogen were assessed in arterial blood before, during and after cardiopulmonary bypass (CPB). Oxygenation and ventilation indices before and after surgery, post-operative chest tube drainage and transfusion volume as well as the duration of assisted ventilation were also measured. A platelet count and fibrinogen dropped less in the Bioline group significantly compared with that in control at the end of CPB (p

Еще

Cardiopulmonary bypass, biocompatible coatings, inflammatory response, respiratory indices