Optimization of parameters of magneto-resonant tomograph

Бесплатный доступ

The problem of increase in a signal providing process of internal organs visualization in the magneto-resonant tomograph is investigated on the basis of the some its parameters optimum choice. The principle of the tomograph work is analysed. On the basis of quantum-mechanical approach the interrelation between an angle of the magnetic moment precession of hydrogen nucleus in an organism, frequency of the variable magnetic field stimulating the precession and size of the used constant magnetic field is found. The given interrelation allows find the optimal parameters of the tomography work.

Hydrogen-containing tissues of organism, magnetic moment, resonant frequency, an angle of precession, proton, magnetic field, magneto-resonant tomograph