The development of analytical methods of validation of the medicalxdrug equipment purification
Автор: Kondratyeva M.L., Naprienko E.N., Shkuratova O.V.
Журнал: Сибирский журнал клинической и экспериментальной медицины @cardiotomsk
Рубрика: Организация производства медицинских препаратов
Статья в выпуске: 2-2 т.26, 2011 года.
Бесплатный доступ
The aim of the research was to develop analytical methods of estimating residual amounts of substances in the equipment purification, and to check sensitivity of these methods. The following methodical approaches were used: reactions for chlorides, sulphates, phosphates, protein, analgene, formaldehyde as well as spectrophotometric and titrimetric methods. In the course of the research there have been developed methods of estimating residual concentration of a wide range of substances in washing waters after purification of the technical equipment with the sensitivity of 1 to 10 ppm (1-10 mcg/ml), the standard purity criterion being 10 ppm.
The validation of purification, residual concentration estimation, sensitivity of methods
Короткий адрес: https://sciup.org/14919496